Home Corporate Products  
   
 
[KCH806] life-support machine
Entire machine outward appearance
The graph only supplies the reference

Outline

KCH806 life-support machine is a usable life-support machine that developed by the unit Zibo Kechuang electronic limited company and Shandong University of Science and Technology in 2002. This machine is air operated, the microcomputer control, the constant volume life-support machine, this machine mainly has function and so on IPPV, SIMV, CPAP, SIGH, PSV, PEEP, may through establish the moist spirit, the breath frequency, the inspiration time, holds one's breath parameter realization and so on the time to patient's breath management, this machine major component uses the high grade import product, enhanced this machine reliability and the stability greatly, this machine structure reasonable, the ease of operation, the performance are reliable, is suitable for the emergency medical treatment, the ICU hospital ward and so on the breath rescue and the treatment.

Main performance and technical specification

  1. Ventilates the pattern: IPPV, SIMVf/2, SIMVf/4, CPAP1, CPAP2, SIGH, PSV, PEEP.
  2. Air course performance:
    A) air course most Da'an full pressure:Is not bigger than 6KPa.
    B) 21% ~ 100%.
  3. Ventilates the performance:
    A) breathes the frequency: 6 ~ 60/minutes, error ±15%.
    B) inspires the time: 0.2 ~ 5 seconds, error ±15%.
    C) holds one's breath the time: 0 ~ 0.5 seconds, error ±15%.
    D) triggers the pressure:The inspiration triggers the pressure in -2KPa ~ -1Kpa, the allowable deviation ±20%; The inspiration triggers the pressure in -1KPa ~ -0.4Kpa, allowable deviation ±200Pa; The inspiration triggers the pressure in -0.4KPa ~ 1Kpa, the allowable deviation ±50Pa, the inspiration triggers the pressure in 1KPa ~ 2Kpa, the allowable deviation ±20%.
    E) moist spirit: 300 ~ 1000ml; Error ±20%.
    F) the minute passes the spirit: Is not smaller than 18 L/min.
    G) PSV pressure: 2 ~ 5 KPa, error ±20%.
    H) PEEP pressure: 0 ~ 2 KPa, error ±20%.
  4. Reports to the police the performance:
    A) the pressure upper limit reports to the police: 2 ~ 6KPa, the error ±20%, the acousto-optics reports to the police.
    B) the low pressure limit reports to the police: 0 ~ 2KPa, the acousto-optics reports to the police. 0 ~ 0.5KPa, error ±100Pa; 0.5 ~ 2KPa, error ±20%.
    C) static sound:Is not bigger than 2 minutes.
  5. Monitor performance:
    A) moist spirit: 0 ~ 1200ml. 0 ~ 300ml, error ±60ml; 300 ~ 1200ml error ±20%.
    B) actual frequency.
    C) passes the spirit.
    D) attracts shouts the ratio.
    E) gas channel pressure: Pressure gauge demonstration.
  6. Entire machine performance:
    A) power source:Exchange 220V±22V.50Hz±1Hz.
    B) main engine power: 30VA.
    C) electrical security: Conforms to GB9706.1-1995 "medical electrical equipment first part: Safe general request "to I kind of B equipment correlation request.
    D) entire machine noise: Is not bigger than 65dB (A).
    E) system compliance: Is not bigger than 4ml/100Pa.
  7. Exploitation conditions:
    A) ambient temperature: +41 ℉ ~ +104 ℉.
    B) relative humidity: Is smaller than 85%.
    C) atmospheric pressure: 96 ~ 104KPa.
    D) uses is mad the source: Oxygen, 0.25 ~ 0.3Mpa, compressed air: 0.25 ~ 0.3Mpa.
    E) preheat interval: 5 minutes.
  8. External dimensions: 470×400×285mm.
  9. Net weight: 16kg.

Zibo Kechuang electronic Co.Ltd & Zibo Kechuang medical instrument Co.Ltd 2005
Shandong ICP prepares 05035933