Entire machine outward appearance
The graph only supplies the reference
Outline
KCH606 series life-support machine is a usable life-support machine that developed by the unit Zibo Kechuang electronic limited company and Shandong University of Science and Technology in 2000. This machine mainly has function and so on IPPV, SIMV, CPAP, SIGH, PEEP, may through the hypothesis breath frequency, attract shouts compared to and adjusts the moist spirit realization to patient's breath management, this machine major component uses the high grade import product, enhanced this machine reliability and the stability greatly, this machine structure reasonable, the ease of operation, the performance are reliable.
KCH606 series life-support machine for microcomputer control, air operated electric control, time cut, limiting pressure life-support machine. The kneading board uses touches the pressed key initialization parameter, and has the indicating lamp demonstration breath pattern; The pressure gauge may monitor the gas channel pressure as necessary, the synchronization forces to breathe the function to be possible the interconversion. The major component uses the high grade import product, does not have the machinery to wear, the work is reliable. Is suitable in the emergency medical treatment, the nerve the surgical department, the operating room, the ICU hospital ward and so on.
This series product attains many item of national monopoly, Patent number: 01261459.9, 02212225.7, 02212224.9
Main performance and technical specification
- Ventilates the pattern:
IPPV (intermittent positive pressure ventilates)
SIMVf/2 (synchronized intermittence control ventilates)
SIMVf/4 (synchronized intermittence control ventilates)
CPAP1 (continues positive pressure to ventilate)
CPAP2 (continues positive pressure to ventilate)
SIGH (sigh breath)
PEEP (at the end of breath positive pressure)
- Air course performance:
A) air course most Da'an full pressure:Is not bigger than 6KPa.
B) inspires the oxygen concentration: 40%th, 60%, 80%, 100%.
- Ventilates the performance:
A) breathes the frequency: 6 ~ 60/minutes, error ±15%.
B) attracts shouts the ratio: 1:1, 1:2, 1:3, 1:4, 1:5, 2:1, 2:3, 3:5, error ±15%.
C) triggers the pressure:The inspiration triggers the pressure in -2KPa ~ -1KPa, the allowable
deviation ±20%; The inspiration triggers the pressure in -1KPa ~ -0.4KPa, allowable deviation ±200Pa; The inspiration triggers the
pressure in -0.4KPa ~ -0.1C$kPa, allowable deviation ±50Pa.
D) moist spirit: 300 ~ 1000ml; Error ±20%.
E) the minute passes the spirit: Is not smaller than 18 L/min.
- Reports to the police the performance:
A) the pressure upper limit reports to the police: 2 ~ 5KPa, the error ±20%, the acousto-optics reports to the police.
B) the low pressure limit reports to the police: 0 ~ 2KPa, the acousto-optics reports to the police. 0 ~ 0.5KPa, error ±100Pa; 0.5 ~ 2KPa, error ±20%.
C) static sound:Is smaller than 2 minutes.
- Monitor performance:
A) moist spirit: 0 ~ 1200ml. 0 ~ 300ml, error ±60ml; 300 ~ 1200ml,error ±20%.
B) actual frequency: The life-support machine each minute supplies patient's actual breathing rate and demonstrated with the digitaltube.
C) gas channel pressure: The patient gas channel actual pressure and instructed with the pressure gauge.
- Entire machine performance:
A) power source:Exchange 220V±22V.50Hz±1Hz.
B) main engine power: 30VA.
C) electrical security: Conforms to GB9706.1-1995 "medical electrical equipment first part: Safe general request" to I kind of B equipment correlation request.
D) entire machine noise: Is not bigger than 65dB (A).
E) system compliance: Is not bigger than 4ml/100Pa.
- Exploitation conditions:
A) ambient temperature: +5 ℃ ~ +40 ℃.
B) relative humidity: Is smaller than 85%.
C) atmospheric pressure: 96 ~ 104KPa.
D) uses is mad the source: Oxygen, 0.2 ~ 0.3MPa.
E) preheat interval: 5 minutes.
- External dimensions: 370×280×260mm.
- Net weight: 10kg.